BL4-2 Hardware Overview
The beamline BL4-2 at SSRL is a dedicated experimental station for small-angle x-ray scattering and diffraction (SAXS) experiments. It focuses on the application of SAXS in the areas of structural biology and biophysics and provides state-of-the-art facilities for a large variety of different experiments. The end station of the beamline is laid out as a regular pinhole SAXS camera with an evacuated flight path of variable length in order to optimize the configuration for different types of experiments.

The beamline uses the central part of the radiation fan originating from a 20-pole, 2 T wiggler as its source. A bent cylindrical mirror with Rhodium coating at mid-point between the source and the detector provides a focused 1:1 image of the source at the detector plane, with a native beamsize of 0.18 mm x 0.8 mm (FWHM). The liquid nitrogen cooled monochromator provides a choice of either a pair of Si111 crystals for high energy resolution or a pair of multilayer diffractive elements for high flux applications with a more relaxed energy resolution. The SAXS camera operates with a set of beam defining/collimating slits and two sets of scatterless guard slits to collimate and shape the beam according to the specific experimental needs. The evacuated flight tube has a choice of seven different sample-to-detector distances that can be semi-automatically changed and houses at the downstream end a small beamstop with integrated photodiode for measuring the transmitted photon flux needed for correctly scaling the data. A photon counting Dectris Pilatus3 X 1M pixel array detector is used for data collection. Alternatively a Rayonix225 HE CCD detector can be employed for specialized applications where smaller pixel sizes and gap-less data is important.
| Source Type | Insertion Device, Wiggler |
|---|---|
| Optics | focusing bent cylinder mirror, Rh coated (cutoff energy 17.5keV) upward reflecting, Magnification 1 |
| Monochromator | Si111 Mo/B4C Multilayer |
| energy range | 6 - 17 keV (for Si111) 8 - 14 keV (for Multilayer) |
| energy resolution | 5 x 10-4 (for Si111) 3 x 10-2 (for Multilayer) |
| flux (in full beam) | 3 x 1012 ph/s (for Si111) 1 x 1014 ph/s (for Multilayer) |
| spot size at detector (FWHM) | (0.18 x 0.8) mm2 |
| Detectors | Dectris Pilatus3 X 1M Eiger 1M |
| sample-detector-distance | 0.25 - 3.5 m, variable in seven discrete steps |
| x-ray source-detector-distance | 32 m |
| typical q - range [Å-1] (at 11keV) | 0.045 - 3.6 (for 0.25m) 0.024 - 2.2 (for 0.5m) 0.016 - 1.7 (for 0.75m) 0.010 - 1.0 (for 1.1m) 0.0068 - 0.67 (for 1.7m) 0.0045 - 0.46 (for 2.5m) 0.0034 - 0.33 (for 3.4m) |