Structural Molecular Biology
Illuminating Biological Structures at the Atomic and Molecular Levels


Biological Small Angle X-ray Scattering/Diffraction
SMB-SAXS Education & Training
SMB-SAXS Data Analysis
Data Analysis Primer
1. Initial Analysis and Quality Assessment of Solution Scattering Data
The BL4-2 pipeline program, SAXSPipe, reduces the 2-D detector images to 1-D intensity vs q curves, I(q), using the program SASTOOL. It runs initial analyses and quality checks on your data. The results are summarized in an HTML report that you can open in any web browser.
SASTOOL produces four types of output files, which are useful for further inspection:
| File | Contents |
|---|---|
| .dat | Scaled, radially integrated 1D intensity curve from a single image |
| .tot | Average of the individual .dat files for one sample. Used to make the .sub file. |
| .sub | Background-subtracted curve (e.g. protein.tot − buffer.tot). This is the file usually used for further analysis. |
| .log | Processing log |
The .dat, .tot and .sub files each have three columns:
- Either the distance from the beam center in pixels or the scattering vector q, depending on the setting of the
-qparameter. - The scattered intensity, I(q).
- The error, which is the smaller of the standard deviation and the square root of the intensity.
Always check which units of q your files use (Å⁻¹ or nm⁻¹). Mixing units between programs is a common source of errors: values such as Rg and Dmax will come out 10 times too large or too small.
2. Checking for Radiation Damage
Before averaging, compare the individual exposures (.dat files) of each sample. In the absence of radiation damage, all frames should overlap within the noise. A gradual increase in intensity at low q from frame to frame is a typical sign of radiation-induced aggregation. Leave damaged frames out of the average. Typically SAXSPipe/SASTOOL will do a reasonable good job in sorting out "bad" data and rejecting it from the overall average. But it is always a good idea to double check, especially if severe radiation damage is observed.
If damage is a problem, these measures can help:
- shorter or fewer exposures
- lower sample concentration
- flowing or oscillating the sample during exposure
- adding radical scavengers such as 2–5% glycerol, DTT or TCEP
Discuss these options with beamline staff.
Programs used: SAXSPipe, Graph-it (available at BL4-2), PRIMUS, BioXTAS RAW.
3. Confirming Buffer Subtraction
Correct buffer subtraction is the starting point for all further analysis. A mismatched buffer or an error in scaling or normalization will give an incorrect subtracted curve. The SAXSPipe report makes this easy to check quickly.
Open the sample and buffer .tot files (the averaged curves before subtraction). At high q, the sample curve should be slightly but consistently above the buffer curve.
- Over-subtraction (buffer scattering higher than sample at high q) produces negative intensities at high q. On a log scale, these show up as a sharp drop or gaps in the curve, because most software can't plot negative values on a log scale.
- Under-subtraction (buffer mismatch or a missing buffer component) leaves a flat, constant background at high q. That background distorts the Kratky plot and the Porod volume.
The best buffer is the final dialysis buffer or the size-exclusion column buffer, measured in the same sample cell under identical conditions.
Programs used: SAXSPipe (available at BL4-2), Graph-it (available at BL4-2), PRIMUS, BioXTAS RAW and other mathematical or graphing software.
4. Guinier Plot
A Guinier plot gives the overall size of the particle as the radius of gyration (Rg), together with the forward-scattering intensity, I(0). I(0) is proportional to the molecular weight times the mass concentration. Equivalently, for a given number concentration of particles, I(0) is proportional to the square of the molecular weight. Both Rg and I(0) are sensitive indicators of whether the sample is behaving well in the X-ray beam.
The plot itself is simple: ln I(q) against q². The plot must be linear over the fitted range. For globular proteins, that range should satisfy qmax · Rg ≤ 1.3, where qmax is the upper end of the Guinier fit. For elongated particles, use a lower limit (about 0.8–1.0). Use the lowest-q data points available, but exclude any points affected by the beamstop.
If the plot isn't linear, there are two typical cases:
- Upturn at low q: aggregation, radiation damage, attractive interparticle interactions, or polydispersity of the sample.
- Downturn at low q: repulsive interparticle interactions, typical of charged proteins at low ionic strength or high concentration.
For a concentration series, check that:
- all concentrations give the same Rg within error, and
- I(0) divided by concentration is the same for all samples.
If Rg or I(0)/c changes systematically with concentration, the data need to be extrapolated to zero concentration (see Section 7).
Programs used: SAXSPipe, Graphit (available at BL4-2), PRIMUS (AUTORG), BioXTAS RAW and other mathematical or graphing software.
5. Kratky Plot
How the intensity decays beyond the Guinier region (the Porod region) reflects the shape and compactness of the particle:
| Power law, I(q) ∝ | Typical structure |
|---|---|
| q⁻⁴ | Compact particle with a sharp surface (Porod's law), e.g. a well-folded globular protein |
| q⁻² | Thin flat disk or sheet, at intermediate q |
| q⁻² | Gaussian chain (ideal random coil) |
| q⁻⁵ᐟ³ | Swollen chain (random coil with excluded volume), e.g. an unfolded protein in good solvent |
| q⁻¹ | Thin rod |
The Kratky plot, I(q)·q² against q, is therefore a useful way to check how globular or flexible a protein is:
- Well-folded, globular protein: a bell-shaped peak at low q, then a decay toward the q axis at higher q.
- Multi-domain protein: additional peaks or shoulders may appear, but the curve still decays at high q.
- Flexible or partially unfolded protein: the curve levels off at a plateau or keeps rising at high q instead of returning toward zero.
The dimensionless Kratky plot, (q·Rg)² · I(q)/I(0) against q·Rg, normalizes for size and concentration, so different samples can be compared directly. For a compact globular particle, the peak lies at q·Rg = √3 ≈ 1.73, with a height of about 1.1. A peak shifted up and to the right indicates elongation or flexibility.
If an atomic or homology model is available, always compare its calculated Kratky plot with the data.
Check that the peak position is the same for all concentrations. Strong interparticle effects, aggregation, oligomerization or dissociation change the curve not only at low q but also in the Porod region. If the shape of the curve at intermediate q changes with concentration, the dominant species in solution is not the same at all concentrations.
Programs used: Graph-it (available at BL4-2), PRIMUS, BioXTAS RAW and other mathematical or graphing software.
6. P(r) Function
The pair-distance distribution function, P(r), is a histogram of all distances between pairs of electrons within the particle. It is obtained from the scattering curve by an indirect Fourier transform. P(r) gives an intuitive real-space picture of the particle's shape:
- a symmetric bell shape indicates a compact globular particle
- a long tail indicates an elongated particle
- several maxima indicate separated domains or subunits
The most important parameter from P(r) is the maximum particle dimension, Dmax: the distance at which P(r) returns smoothly to zero. Dmax is usually found by testing several values and choosing the one that gives a smooth curve, which approaches zero gradually rather than being forced to zero or oscillating.
To keep errors in the Fourier transform small, the data need good coverage at both low and high angles. In particular, the lowest measured q should satisfy qmin < π/Dmax.
The real-space Rg calculated from P(r) should agree with the Guinier Rg. The real-space value uses the whole scattering curve, so it is less sensitive to noise or a few bad points at the lowest angles. A significantly larger real-space Rg often points to aggregation or a poorly chosen Dmax.
P(r) can also be calculated for flexible proteins. However, it is then a weighted average over many conformations, and Dmax is harder to define. In that case, use P(r) descriptively, and avoid single-shape modeling (see Section 8).
Programs used: GNOM (via the PRIMUS interface or DATGNOM), BioXTAS RAW (GNOM and BIFT).
7. Molecular Weight Estimation
There are several ways to estimate the molecular weight (Mw) of your sample, each with its own assumptions and errors. Cross-validating several methods is very important. Agreement between independent estimates within about 10–20% is a good sign. Larger discrepancies point to problems with concentration, buffer subtraction or sample quality.
A. Absolute scale from water scattering
The scattering of pure water is known and independent of temperature to first order, so it can be used to put I(0) on an absolute scale. Mw then follows from I(0), the sample concentration, the partial specific volume (about 0.73 cm³/g for proteins) and the scattering contrast. The program Graphit at BL4-2 performs this calculation, and beamline staff provide water scattering data during setup. See Orthaber, Bergmann & Glatter, J. Appl. Cryst. (2000) 33, 218–225.
B. Comparison with a protein standard
Measure a standard protein of known Mw and concentration under the same conditions. Then:
Mw,sample ≈ Mw,standard × [I(0)sample / csample] ÷ [I(0)standard / cstandard]
where c is the mass concentration (e.g. mg/mL). The standard should be a monodisperse, globular protein in an appropriate buffer, such as lysozyme, RNase A or glucose isomerase. Do not use BSA, which typically contains a mixture of monomers, dimers and higher oligomers. Note that this method assumes the sample and standard have similar contrast (for example, a protein sample compared with a protein standard). It is not reliable for nucleic acids or glycosylated or lipid-containing samples.
C. Porod volume
The Porod volume (Vp) is the volume of the hydrated particle, and it doesn't depend on concentration. For globular proteins:
Mw (kDa) ≈ Vp (nm³) / 1.6, or equivalently Mw (Da) ≈ 0.625 × Vp (ų)
PRIMUS (through its GNOM interface) can calculate the Porod volume alongside P(r). The Porod volume depends strongly on correct buffer subtraction and on the q range used, and it becomes unreliable for flexible proteins. See Petoukhov et al., J. Appl. Cryst. (2012) 45, 342–350.
D. Volume of correlation (Vc)
The volume of correlation is a concentration-independent invariant that also works for flexible proteins and RNA. See Rambo & Tainer, Nature (2013) 496, 477–481. It is available in ATSAS (DATMW) and BioXTAS RAW.
E. Dummy-atom models
If you have already run ab initio modeling with DAMMIF or DAMMIN, Mw can be estimated from the volume of the dummy-atom model:
Mw (Da) ≈ VDAM (ų) / 2
F. SAXSMoW and Bayesian estimates
SAXSMoW (Fischer et al., J. Appl. Cryst. (2010) 43, 101–109) estimates Mw from a single curve on a relative scale, using a corrected Porod invariant. ATSAS DATMW and BioXTAS RAW also include a Bayesian method that combines several of these estimates into a single value with a credibility interval.
8. Concentration Effects and Merging Curves
Interparticle interactions make the scattering curve depend on concentration, mainly at low q:
- Repulsion lowers I(0) and Rg at higher concentrations.
- Attraction or aggregation raises them.
If Rg, I(0)/c or the low-q shape changes with concentration, extrapolate the data to infinite dilution:
- Normalize each curve by its concentration.
- Extrapolate the intensity at each q value linearly to zero concentration.
- Confirm that the lowest-concentration curves no longer change.
Low-concentration data are reliable at low q, but noisy at high q. High-concentration data have better signal at high q, where interparticle effects are negligible. You can therefore combine the two: use the zero-concentration (or lowest-concentration) curve at low q and the highest-concentration curve at high q. Before merging:
- make sure both curves are scaled to the same concentration
- choose an overlap region where both curves agree
- confirm that the Kratky peak position is the same in both curves, so the particle shape is the same at both concentrations
Programs used: Graphit (available at BL4-2), PRIMUS, BioXTAS RAW and other mathematical or graphing software.
9. Modeling and Further Analysis
A wide range of programs is available for interpreting SAXS data in terms of 3D structure. Choose the approach that matches what your initial analysis showed:
Comparing with known atomic structures. If a crystal structure, cryo-EM structure, NMR structure or prediction (such as an AlphaFold model) is available, calculate its theoretical scattering curve and compare it with the data, using CRYSOL, FoXS or Pepsi-SAXS. The fit quality is usually reported as χ², and a value near 1 indicates good agreement. Systematic deviations often reveal missing flexible regions, a different oligomeric state, or a conformation in solution that differs from the one in the crystal.
Ab initio shape reconstruction. DAMMIF, DAMMIN and GASBOR build low-resolution models from beads or dummy residues that reproduce the scattering curve. Run multiple independent reconstructions (typically 10–20), and compare and average them with DAMAVER or DAMCLUST to judge how unique the solution is. These methods assume a single compact, rigid particle. They should not be used for flexible or polydisperse samples.
Rigid-body and hybrid modeling. When structures of individual domains or subunits are known, programs such as SASREF, CORAL and BUNCH find their arrangement, and can model missing linkers. SREFLEX refines high-resolution models with normal-mode analysis to match the solution data.
Flexible systems. If the Kratky plot shows flexibility, describe the protein as an ensemble of conformations rather than a single shape. Programs include EOM (Ensemble Optimization Method), MultiFoXS and BILBOMD.
Mixtures. For samples that contain several species, such as a monomer–dimer equilibrium or a partially formed complex, OLIGOMER can estimate the volume fraction of each component from their known structures. For SEC-SAXS data, evolving factor analysis (EFA) and related methods in BioXTAS RAW, as well as CHROMIXS in ATSAS, can separate overlapping elution peaks.
Most standard programs assume a monodisperse, single-component system without flexibility. Make sure your initial analysis supports that assumption before modeling. Always validate models against the experimental data and, where possible, against independent information from other methods.
10. Reporting and Data Deposition
Community publication guidelines for biomolecular SAXS recommend reporting the following:
- sample details
- data collection parameters
- the structural parameters (Rg, Dmax, Mw) and the method used for each
- the software used
- model-fitting statistics
See Trewhella et al., Acta Cryst. D (2017) 73, 710–728. We strongly encourage depositing your data and models in the Small Angle Scattering Biological Data Bank (SASBDB) at sasbdb.org. Many journals expect this, and it makes your results easier to reproduce and reuse.