BL9-3

    Overview

    Beam line 9-3 is a high flux focused wiggler side-station optimized for x-ray absorption spectroscopy (XANES and EXAFS) in the 5.5–30 keV range, dedicated for x-ray absorption spectroscopy and EXAFS measurements on dilute biological systems. A dedicated LHe cryostat allows for routine low temperature measurements. The station features a double-crystal Si(220) monochromator with energy resolution up to 10-4(ΔE/E) and small vertical entrance slits to reduce intensity glitches. Equipment includes a Canberra 100-element Ge detector for dilute samples, a Mirion 24-element Ge detector, a passivated implanted planar silicon (PIPS) semiconductor detector for fluorescence detection, ionization chambers for transmission data, and a LHe cryostat for maintaining samples at ~10 K (4-200 K with LHe dewar).

    Source: 20-pole, 2-Tesla wiggler side station, 0.8 mrad acceptance

    Status

    Operational

    Beamline Parameters

    Energy Range Resolution ΔE/E Spot Size Flux
    Focused 5500-30000 eV 1 x 10-4 1 x 4 mm2 ~2 x 1012

    Optics

    M0 Mirror M1 Mirror Monochromator
    Vertically collimating, flat, bent, 1 m, Si, Rh-coated Cylindrical, bent, 1.2 m, Zerodur, Rh-coated Si(220), ϕ = 0o & 90o. double-crystal, variable-exit geometry

    Source

    Wiggler

    Instrumentation

    Detectors Canberra 100-element germanium solid state detector. Mirion 24-element germanium solid state detector with Xspress3X electronics. Passivated implanted planar silicon (PIPS) semiconductor detector. Ionization chambers.
    Ancillary Soller slits and scatter guards, sample shutter to alleviate beam damage, automatic scan queueing for multiple samples/multiple spots on samples. Oxford Helium cryostat, cooled by liquid He dewar (4-200 K). Multiple-sample holder for room temperature samples.

    Sample Environment

    Standard room temp Ambient pressure Single sample holder 4 slot sample holder 8 slot sample holder
    Standard LHe cryostat Sample temperature ~10 K: Single solid sample 4-slot solid sample cassette Ssingle solution sample 5-slot solution sample cassette
    Non-standard setups Communicate with beam line staff well in advance to ensure your set-up will work.

    Absorption

    768 microns Be, 205 microns C


    SPEAR3

    Publications

    placeholder

    Science Highlights