BL5-2

    Overview

    BL5-2 is a plane grating monochromator beam line optimized for high-resolution, micro-focused angle-resolved photoemission spectroscopy (ARPES) in the photon energy range of18 - 200 eV with full polarization control. The dedicated experimental end station is equipped with a SCIENTA DFS30 electron spectrometer. This system is capable of performing ARPES experiments with a total energy resolution of a few meV and an angular resolution of ~0.1°. It is also equipped with a low-temperature sample manipulator with three degrees of freedom of rotation, which can be cooled down to ~7 K.

    Because BL5-2 and BL5-4 cannot operate simultaneously, they are often shown combined as BL5-2/BL5-4. BL5-2 is the default station for majority of the users on BL5. Please communicate with the responsible beam line scientists in advance if BL5-4 is the desired experimental station.

    Source: 15-period undulator, 31-pole APPLE-II EPU, 0.65 mrad acceptance.

    Status

    Operational

    Beamline Parameters

    Energy Range Resolution ΔE/E Spot Size Flux
    Focused 18 - 200 eV ≥ 2.5 X 10-5 0.036(H) X 0.006(V) mm2 ~ 2 X 1012
    *@500 mA, 50 eV, E/ΔE = 10,000

    Optics

    M0 Mirror M1 Mirror M2 Mirror Monochromator M3 Mirror
    horizontal deflecting, flat, water-cooled horizontal deflecting, flat, water-cooled vertical deflecting, flat, water-cooled Variable Line Spacing Plane Grating Monochromator (VLSPGM)
    • 1200 l/mm, c = 1.5 (HOSG)
    • 1200 l/mm, c = 3.0 (HRG)
    • 600 l/mm, c = 4.0 (HRG2) (planned)
    Horizontal and vertical refocusing, elliptical toroidal

    Source

    Undulator

    Instrumentation

    Detector SCIENTA DFS30 Electron Analyzer
    Manipulator Low-temperature 6-axis Manipulator
    Alkaline metal doser (K, Rb, Li)

    Sample Environment

    UHV < 3 X 10-11 Torr
    Sample temperature < 7 - 300 K
    Sample size < 10 X 10 mm2
    Annealing temperature < 1000 °C


    SPEAR3

    Publications