Overview
Beamline 15-2 is an undulator experimental station dedicated to advanced hard x-ray spectroscopy. The end station is equipped with sub-corehole lifetime resolution x-ray emission spectrometers (a 7-crystal Johann spectrometer and a 8 crystal von Hamos instrument) as well as a high-throughput and high energy resolution (0.3eV) 40-crystal low-q x-ray Raman spectrometer. The end-station also incorporates a high-power MHz repetition laser together with fast diodes and FPGA electronics enabling laser pump x-ray probe time-resolved spectroscopies (~75ps time resolution) using the camshaft bunch (~20mA) of SPEAR3. Advanced sample environments including electrochemical cells, thermal catalysis reactors, cryostats, continuous flow liquid jets, microfluidic mixers and other in situ environments enable a multipurpose scientific program across catalysis, energy science, biochemistry, environmental science as well as actinides research.
Status —
Beamline Parameters —
Energy Range | Resolution ΔE/E | Spot Size | Flux | Angular Acceptance | |
---|---|---|---|---|---|
Focused | 4800-18200 | 1.0-15 10-5 | 5x35um2 (or 100x700um2) | ~3x1013 ph/s (Si111) ~ 4x1012 ph/s (Si311) |
N/A |
Unfocused | 18000-24000 | 10-4 | 500x2000um2 | ~ 4x1011 ph/s (Si311) | N/A |
Optics —
M1 KB mirrors: Bent single crystal KB mirrors for a vertical and horizontal beam focusing. Rh and Si stripes provide selectable cutoffs at 11keV and 18.2 keV, respectively. |
---|
Monochromator: LN2 cooled, Si(111), ϕ = 0o , Si(311), ϕ = 0o. double-crystal. |
Source —
Instrumentation —
Detectors | Multiple single element Vortex solid-state; PIPS diodes, APD diodes, ionization chambers, Pilatus, and Eiger-X. |
---|---|
Cryostat | Oxford LHe cryostat (4-200 K) |
Other | Integrated fast shutter system and variable filters to alleviate radiation damage. High power and high rep rate laser: Amplitude Tangerine 50W Colby time-delay unit Teledyne Wavemaster 816ZI-B oscilloscope 16 GHz, 80 GS/s) Teledyne ADQ-14bit 4-channel digitizer card (1GSPS/channel). |
Spectrometer | 7-crystal XES spectrometer (http://dx.doi.org/10.1063/1.4803669) |
40-crystal XRS (http://dx.doi.org/10.1063/1.4704458) | |
8-crystal von-hamos XES spectrometer (https://doi.org/10.1063/1.4737630) |
Sample Environment —
Ambient pressure experiments, He chamber(s), flexibility with various in-situ setups including electrochemical cells and thermal reactors for operando catalysis, liquid samples and liquid jets, high pressure. Dedicated LHe cryostat (4-200 K). |
---|
Minimum sample requirement (in terms of concentration): Typically, end-station can measure sub-100umol concentrations and sub monolayer materials. Feasibility depends strongly on the technique and the scientific question. Please contact responsible scientific staff for guidance. |
Absorption —
127um Be |
---|