BL15-2

    Overview

    Beamline 15-2 is a state-of-the-art undulator experimental station specializing in advanced hard x-ray spectroscopy. The end station features cutting-edge instrumentation, including a 7-crystal Johann spectrometer and an 8-crystal von Hamos instrument for sub-core hole lifetime resolution x-ray emission spectroscopy. It also houses high-throughput and high-energy-resolution x-ray Raman spectrometers with resolutions of 0.3 eV, optimized for both low-q (40-crystal) and high-q (30-crystal) scattering geometries. For time-resolved studies, the beamline integrates a high-power MHz repetition rate fiber laser with fast diodes and FPGA electronics, enabling laser-pump x-ray-probe spectroscopy with ~75 ps time resolution, utilizing SPEAR’s camshaft bunch (~20 mA). The versatile end-station supports a wide range of advanced sample environments, including electrochemical cells, thermal catalysis reactors, cryostats, continuous-flow liquid jets, and microfluidic mixers. These capabilities enable a robust and diverse scientific program spanning catalysis, energy science, biochemistry, environmental science, and actinides research.

    Source: Undulator (87periods with 22mm/period), low emittance (6.2nrad) lattice.

    Status

    Operational

    Beamline Parameters

    Energy Range Resolution ΔE/E Spot Size Flux Angular Acceptance
    Focused 4800-18200 1.0-15 10-5 5x35um2 (or 100x700um2) ~3x1013 ph/s (Si111)
    ~ 4x1012 ph/s (Si311)
    N/A
    Unfocused 18000-24000 10-4 500x2000um2 ~ 4x1011 ph/s (Si311) N/A

    Optics

    Monochromator: Mirrors
    LN2-cooled Si(111), ϕ = 0o. double-crystal, fixed-exit geometry LN2-cooled Si(311) , ϕ = 0o. double-crystal, fixed-exit geometry No mirror before mono. A KB mirror system after the mono: KB-M1 mirror (V): Rh-coated Si mirror; 500mm long. KB-M2 mirror (H): Si and Rh coated stripes; 800mm long.

    Source

    Undulator

    Instrumentation

    Detectors Multiple single element Vortex solid-state; PIPS diodes, APD diodes, ionization chambers, Pilatus, and Eiger-X.
    Ancillary Integrated fast shutter system and variable Al filters to alleviate radiation damage. High power and high rep rate laser: Amplitude Tangerine 50W, Colby time-delay unit, Teledyne Wavemaster 816ZI-B oscilloscope (16 GHz, 80 GS/s), Teledyne ADQ-14bit 4-channel digitizer card (1GSPS/channel), Biologic potentiostat, Asahi Xe light source.
    Spectrometer 7-crystal XES spectrometer (http://dx.doi.org/10.1063/1.4803669)
    40-crystal XRS (http://dx.doi.org/10.1063/1.4704458)
    8-crystal von-hamos XES spectrometer (https://doi.org/10.1063/1.4737630)

    Sample Environment

    Standard room temp Flexible ambient pressure experiments in both air and under controlled He/N2 chamber(s). Liquid cells and replenishing liquid jets.
    Standard LHe cryostat Sample temperature ~10 K: single solid sample 4-slot solid sample cassette single solution sample 5-slot solution sample cassette
    Non-standard setups Flexibility with various in-situ setups including photo/electro-chemical cells and thermal reactors for operando catalysis. Other customized and/or Users developed sample environments are regularly being integrated. Please, communicate with beam line staff well in advance to ensure your set-up will work.

    Absorption

    127um Be


    15-2