Guide to XAFS Measurements at SSRL


Beamline Alignment



 Figure 1.  Basic Components

The basic components of a typical XAFS experiment are shown above (Figure 1).  X-rays are produced in the source (bending magnets, wigglers, or undulators) and pass through an aperture before striking the monochromator.  This aperture, a pair of slits, can be adjusted to control the size and energy resolution of the beam.  Most monochromators at SSRL XAFS beam lines are double crystal designs.  The first crystal of the pair can be detuned (rotated off of the Bragg condition) for precise alignment and for harmonic rejection.  After entering the hutch, the beam passes through another defining aperture (Huber Ta slits).  The beam passes through the gas in the gas ionization chambers, and its intensity is measured.  These gas ionization chambers are typically used to measure the incident beam when filled with appropriate gas mixtures.  Lytle or Germanium detectors are then used to measure fluorescent x-rays from the samples.

The first step in an XAFS experiment is to align the beam line so that the most intense part of the beam is directed toward the hutch.  The beam alignment procedure is as follow:


 
Beam Alignment Procedure
1. Move monochromator to relevant energy (typically the value of the mid region of the energy scan).
2. Do a table scan 
3. Check the detuning level and set it to the desired amount.  The table scan (step 2) should be repeated if the beam was found to be  detuned more than 70%. 
4. Remove the Huber Ta slit in front of the first ionization chamber (Io).  (If this slit was mounted, the beam would then have to travel through two different vertical apertures, namely the vertical monochromator slit and the vertical Huber Ta slit, producing a convolution intensity profile)
5. Translate the vertical monochromator slit up and down and observe Io.  The goal in this step is to maximize IoNote:  for most beam lines, the monochromator vertical slit translation control knob is located on the top of the monochromator housing. 
6. Replace the Huber Ta slits back in front of the first ionization chamber.
7. Do a table scan to direct the most intense part of the beam through the Io slits and into Io
8. Do a sample position scan, first vertically and then horizontally, to make sure that the incident beam is not blocked by the sample holder. 
9. Place both the mirror and the fluorescent screen at the end of the last gas ionization chamber.  Observe the beam through the mirror.  If the beam is not centered vertically or horizontally, use the rail motor controller to center the beam. 
*Note:  The beam line alignment should be checked after each fill or orbit shift
 

Gas Ionization Chambers
 
SSRL has two standard sizes of gas ionization chambers: 15 cm and 30 cm in length.  In a typical EXAFS experiment, the first ion chamber should be filled with a gas mixture to absorb about 20 to 40% of the incident beam.  A second ion chamber is placed after the sample to detect the transmitted beam, and its gas mixture is also chosen to absorb about 20 to 40% of the transmitted beam.  An appropriate gas mixture fills the third ion chamber to absorb as much of the remaining beam as is possible.  Table 1 below lists the useful gases, densities, absorption edge energies, and absorption coefficients.  The ratios of beam transmitted for ion chambers of length 15 cm and 30 cm for different energy ranges are also listed.  These ratios can be calculated using the following expression:

 

Ratio of the beam transmitted = I1/Io = e-mrL


where Io is the photon flux of the incident beam, I1 is the photon flux of the transmitted beam, m is the absorption coefficient (cm2/g), r is the density of the gas (g/cc), and L is the length of the gas chamber (cm).  By consideration of m of different gases and the desired percentage of transmission, an appropriate gas or gas mixture for each ion chamber can be chosen.  Such an analysis is presented in Table 1.
 
     SSRL provides argon and nitrogen gases.  Users can order other gases at their own expense.

(

  *

:  close-to-optimum compositions excluding mixing of gases)

 
Gas 
(Density g/cm3
(Absorption Edge eV)
Energy Range 
(keV)
m  
(cm2/g)
I1/I0 
(L = 15 cm)
I1/I0  
(L = 30 cm)
Helium 2 6.317 0.98 0.97
(1.78 x 10-4) 3  1.561  1.0  1.0
 (24.6) 5  0.2611  1.0  1.0
  8  0.04951  1.0  1.0
   10  0.02241  1.0  1.0
   15  0.00530  1.0  1.0
   20  0.001919  1.0  1.0
   30  0.0004558  1.0  1.0
 Nitrogen  2  493  9.7 x 10-5  9.4 x 10-9
 (1.25 x 10-3)  3  148.2  0.062  0.0039
 (401.6)  5  30.66  0.56  0.32
   8  6.853  0.88  0.77
  10  3.324  0.94  0.88
  15  0.8793  0.98  0.97
   20  0.3396  0.99  0.99
   30  0.0884 1.0  1.0
 Neon  2  1258  4.2 x 10-8  1.8 x 10-15
(9.0 x 10-4  3  405.6 0.0042   1.8 x 10-5
(866.9)   5  91.06 0.29  0.086
   8  21.81 0.74  0.55
   10  10.90  0.86   0.75
   15  3.037 0.96  0.92
   20  1.213 0.98  0.97
   30  0.3294 1.0  0.99
 Argon  2  513.5  1.1 x 10-6  1.2 x 10-12
 (1.78 x 10-3)  3  167.4 0.011   1.3 x 10-4
(3202)   5  438.6  8.2 x 10-6   6.8 x 10-11
   8  120.2  0.040  0.0016
   10  63.38  0.18  0.034
   15  19.14  0.60  0.36
   20  8.01  0.81  0.65
   30  2.295  0.94  0.88
 Krypton  2  3870  5.3 x 10-95  2.8 x 10-189
 (3.74 x 10-3)  3  1359  7.8 x 10-34  6.1 x 10-67
 (K-edge = 14.3  5  348.9  3.2 x 10-9  1.0 x 10-17
 LI = 1921.0  8  95.92 0.0046   2.1 x 10-5
 LII = 1727.2  10  51.26 0.0056   0.0031
LIII = 1674.9)   15  115.2 0.0016  2.4 x 10-6 
   20  54.84 0.046   0.0021
   30  18.13  0.36  0.13
Table 1.  Useful gases, densities, absorption coefficients, and ratios of transmission

Gain Settings For Ionization Chambers

The voltage-to-frequency (V/F) converters are most linear between values of 0.5 V and 10 V.  Hence, the gain for each Keithley amplifier for Io, I1, and I2 should be adjusted such that the V/F Converters give results in this range.  The same also applies for Lytle detectors.  Also, when working with Keithley amplifiers, it is essential to remember to set the suppression equal to the inverse of the gain.

The Lytle Detector

The Lytle detector, or the fluorescent ion chamber detector, measures fluorescent x-ray emission.  Fluorescence arises from the de-excitation of electrons into the core hole produced by x-ray photon absorption and is proportional to the probability of x-ray photon absorption.  Hence, fluorescence can be used to measure XAFS.  Fluorescence measurement has the greatest utility when using samples that are dilute, have low K-edge energies, or are very small.  Depending upon the system being studied, it can be effective for concentrations down to one part per million.  Typically, the Lytle dectector is used for samples having greater than 1000 parts per million of the metal ion of interest.  For more dilute samples, it is often necessary to use a germanium detector.  The advantages of the Lytle detector are its much larger solid angle of acceptance and its ability to measure photons at much higher count rates than germanium detectors.  The use of germanium detectors are not covered in this document due to their complexity.


When using a Lytle detector, most samples should be positioned such that the angle of incidence and exit are equal to 45o, which minimizes elastic scattering into the detector and maximizes the signal from the sample.  The standard Lytle detector sample holder provides the 45o geometry.  The sample should be mounted on the back of the holder. Additionally, one must insure that the Soller slits are positioned so that the fan of blades points in the direction of the x-ray beam (refer to Figure 2).  For information on x-ray filters, refer to the section How to Select X-ray Filters.

An important consideration for fluorescence measurements is what gas to use in the ion chamber of the Lytle detector. In general, it is desirable to use nitrogen for x-ray energies below 3 KeV, argon between 3-7 KeV, and krypton (or heavier gases) for energies above 7 KeV.  For greater detail, refer to Table 2. Also, refer to the section Gas Ionization Chambers.  When selecting gases for the Lytle detector, the theory is the same as for ion chambers, except that the Lytle chamber thickness is 3 cm and calculations should be based upon the energy of the fluorescence peak of the element of interest rather than the incident beam.  The gas should absorb 90-95% of the radiation passing through the 3 cm distance of the Lytle chamber.  Therefore, it is undesirable to use too heavy of a gas.

 
Energy (KeV)
8
He
N
Ne
Ar
Kr
Xe
2
m(cm2/g)
L(cm)
6.317 
(887)
493.0  
(1.62)
1258 
(.88)
513.5 
(.92)
3870 
(.059)
2158 
(.079)
3
m(cm2/g)
L(cm)
1.561 
(3589)
148.2 
(5.40)
405.6  
(2.74)
167.4 
(3.35)
1359 
(.12)
799.0 
(.21)
5
m(cm2/g)
L(cm)
.2611 
(21459)
30.66 
(26.1)
91.06 
(12.2)
438.6  
(1.28)
348.9 
(.77)
654 
(.26)
8
m(cm2/g)
L(cm)
.04951 
(114000)
6.853 
(117)
21.81 
(50.9)
120.2 
(4.66)
95.92  
(2.8)
310.5 
(.55)
10
m(cm2/g)
L(cm)
.02241 
(250000)
3.324 
(241)
10.90 
(102)
63.38 
(8.8)
51.26 
(5.2)
170.3  
(1.0)
15
m(cm2/g)
L(cm)
.00530 .8793 
(910)
3.037 
(366)
19.14 
(29)
115.2  
(2.3)
55.96 
(3.0)
20
m(cm2/g)
L(cm)
.001909 .3396 
(2356)
1.213 
(916)
8.01 
(70)
54.84  
(4.9)
24.97 
(6.8)
30
m(cm2/g)
L(cm)
.0004558 .3294 
(9050)
.3294 
(3374)
2.295 
(244)
18.13  
(14.7)
7.82 
(22)
Xe 8 Density (g/cm3) Absorption Edge (eV)
  8 5.9x10-3
34561.4
LI 1921.0
LII 5452.8
LIII 4782.2
 
 * : closest-to-optimal composition (excluding mixing of gases)
Table 2.  Photoelectric Absorption Coefficient, m (cm2/g), and Absorption Length, L (cm)
The absorption length is defined as L, for which m r L=1
 


To adjust the gain on the Lytle detector, simply move the knob mounted on the front of the detector to the desired setting.  For concentrated samples, the typical setting is 1.  For dilute samples, the typical setting is 10 or higher, depending upon the concentration.  Refer to the section Gas Ionization Chambers for gain settings of ion chambers.

Transmission Measurements

In a transmission measurement, the absorption spectrum of the sample is measured directly, as opposed to fluorescence or electron yield experiments.  Key advantages to the transmission method are its simplicity and the absence of self absorption (nonlinear amplitude reduction effects), which is an inherent problem for fluorescence measurements of concentrated samples.  Hence,  transmission measurements are typically used for concentrated samples.  Transmission data are plotted as Ln (Io/I1), which expresses absorption amplitude as a linear function of absorber concentration (and thickness).  For transmission measurements, it is important that the gas ionization chambers be filled with appropriate gas mixtures (see Table 1 in Gas Ionization Chambers for useful gases and percentages of beam transmitted at various energy ranges).  For a typical transmission measurement, the first gas ionization chamber should be filled with a gas mixture that will absorb about 22% of the incident beam.  The second gas ionization chamber should then be filled with a gas mixture that will absorb as much of the remaining beam as possible in order to obtain the best counting statistics.

The following procedure can be used as a guide to preparing transmission samples.  It is based on the assumption that the absorption coefficient (mmetal) of the metal is much larger than the absorption coefficient of the matrix (mmatrix).  If mmetal of the sample is comparable with mmatrix, the matrix will contribute substantially to background noise.  For example, in a sample with two types of metal ions and both present in significant concentration (e.g. PbFe4O7), both metal ions will be dominant absorbers.  Hence, the optimum sample mass may be different from the one obtained using the procedure that follows.

The absorption by a sample with absorption coefficient ms and thickness L is related to the ratio of Io and I1 as exp (-msrL) = (I1/I0).  Consideration of counting statistics in a transmission measurement has shown that the optimum product of  msrL is 2.6, that is when the sample absorbs about 96% of the beam (Stern & Heald (1983), Brown, et. al.).  However, the samples are often prepared in sample holders with standard sizes of 1 or 2 mm.  Therefore, it is practical to achieve this optimum value by adjusting the sample's density instead of its thickness.  In practice, best results have been obtained when the sample is prepared with a density that will absorb about 90% of the beam, that is, when msrx = 2.3.  For a sample with known composition and density, ms can be calculated using the following summation:

 

ms = Si fimi


where fi is the mole fraction of each element i presents in the sample and mi is the absorption coeffient of that element at a particular x-ray energy.  For most cases, the metal will be a dominant absorber.  So, for simplicity, the absorption coeffient of the metal is often used as ms.  Since the effective density of the sample is defined as the mass of the absorber divided by the total volume of the sample, the mass of the absorber can be found using the following expression

 

(msrL) = (msmaL) / Vt = 2.3


where ma is the mass of the absorber and Vt is the total volume of the sample.  Take ms to be equivalent to m of the metal of interest.  Since V and L are fixed, the optimum mass of the metal of interest in the sample can be found.  From this, one can directly calculate the mass of the sample to use.  The sample is then mixed with enough inert matrix to fill the sample holder.
 
It is often necessary to mix the sample with an inert matrix such as BN, B2O3 or LiCO3 to obtain the required sample volume.  A good transmission XAFS sample must be uniform, with the compound of interest grounds to a size smaller than its characteristic absorption length, typically less than 4 mm.

How to Select X-ray Filters 

With the fluorescent method of detection one of the chief concerns is noise generated by the scattering of elastic plus Compton radiation by the sample.  X-ray filters may be used to reduce the scattering radiation that enters the detector, making the choice of an x-ray filter an important consideration.

When selecting a filter, the K-edge (or L-edge) energy of the filter should be such that it absorbs much of the elastic plus Compton radiation scattered by the sample, while simultaneously transmitting the desired Ka radiation.  For elements from titanium (Ti) to ruthenium (Ru), it is preferable to select a x-ray filter with element that has an atomic number of one less than the sample – that is, Z-1.  Beyond Ru, either Z-1 or Z-2 elements may be used.  However, if the L-edge of a sample is be measured, it may be difficult to find an element that satisfies the Z-1 or Z-2 condition for the filter.  For example, consider a uranium sample, whose La, Compton, and elastic peaks are illustrated in Figure 4.  The elements satisfying the Z-1 or Z-2 condition are protactinium and thorium, both of which are unavailable to use as filter elements.  Therefore, the periodic table must be searched to find an element that is both suitable and available.  That is, an element must be found with a K-edge energy that will absorb the Compton and elastic radiation from the uranium sample.  A search reveals that strontium is an excellent candidate, with a K-edge energy that will absorb much of the scattering radiation going towards the detector.
 

Figure 4.  Uranium La, Compton, and Elastic Scattering Peaks

The x-ray filter should be positioned so that it covers the Soller slits.

One consideration when choosing a filter is its thickness.  For dilute samples, it is often preferable to use a filter of 3 absorption lengths.  For more concentrated samples, a filter of 6 absorption lengths probably will be optimal.  However, each should be tried to ascertain which is best.  Included is a table of selected elements to illustrate this (Table 3).

 
Element Density (g/cm3) Mass Absorption Coef. Below and Above K-Edge (cm2/g) Thickness for 3 Absorption Lengths (microns) Transmission Below and Above Edge for 3 Absorption Lengths  
I/I0
Transmission Below and Above Edge for 3 Absorption Lengths  
I/I0
Ti 4.54 93 
685
9.6 .665 
.050
.445 
.0025
V 6.11 77 
612
8.0 .686 
.05
.47 
.0025
Cr 7.19 74 
445
9.4 .606 
.05
.367 
.0025
Mn 7.42 64 
431
9.4 .641 
.05
.410 
.0025
Fe 7.86 60 
397
9.6 .635 
.05
.404 
.0025
Co 8.90 54 
416
8.1 .667 
.05
.459 
.0025
Ni 8.9 47 
290
11.6 .614 
.05
.378 
.0025
Cu 8.94 41 
280
12.0 .644 
.05
.415 
.0025
Zn 7.14 36 
280
15.0 .680 
.05
.462 
.0025
Ga 5.9 31 
231
22.0 .669 
.05
.447 
.0025
Ge 5.32 30 
210
26.8 .651 
.05
.424 
.0025
 Table 3.  Selected Filter Information

Resolution and Slit Dimensions

The energy resolution of an XAFS beamline is controlled primarily by the monochromator and the vertical divergence of the beam.  On many SSRL XAFS beam lines, the vertical beam divergence, described by the vertical opening angle of the beam, is defined by the vertical slits upstream of the monochromator.  The vertical opening angle can be calculated trigonometrically from the source-slit distance and the vertical opening of the slits:

 
Dq = sin-1 (s / d)


where Dq is the angular opening of the slits, s is the monochromator slit dimension, and d is the distance between the source and the monochromator entrance slit.  Distances from sources to monochromator slits for selected beam lines are provided in Table 4 below.
 

Beamline
Source-to-Slit Distance (m)
2-3 16.81
4-1 19.12
4-2 24.01
4-3 17.15
7-2 23.42
7-3 18.47
9-3 16.62
 Table 4.  Distances from the source to the monochromator slit for beam lines at SSRL
 


Once the angular opening is known, the dispersion of the incident beam can be calculated using the differential form of the Bragg's law
 

DE = E cot(qBragg) Dq


where DE is the energy width (FWHM) due to the beam divergence, E is the incident beam energy (eV), qBragg is the Bragg angle for the monochromator and Dq is the angular opening of the monochromator slits.  qBragg can be calculated from Bragg's law, expressed in terms of energy:

 
E = (6.21 x 10 -7 eV•m) / (dsinq) = 12,398.54 eV• A) / l 


For details on energy resolution obtained when focussing mirrors or other optics are inserted into a beam line, see Sean Brennan's "Flux Calculations for SPEAR."

Monochromator Calibration

Energy calibration of monochromators is needed because their calibration drifts over time due to missed steps by the driving motor, computer problems, or changes in the synchrotron orbits.  Therefore, energy calibration should be checked frequently, at least once a day after new fills.  Reference materials (e.g. metal foils that can be checked out from the SSRL store room) are used as energy calibration for the position of the absorption edge.  Absorption edges for other elements should be found near their tabulated energies after this calibration.  The procedure for monochromator calibration is as follow:
 
Monochromator Calibration
1. Place the reference metal foil between Io and I1 or between I1 and I2.
2. Do a scan to collect an edge on the foil using the same resolution in the edge region for the foil as you are using for samples 
3. Plot Ln (Io/I1) or Ln (I1/I2)
4. Check the inflection point of the foil edge.
5. Record the inflection energy. 
6. Move monochromator to the inflection energy. 
7. Reset monochromator to the reference energy of the metal that you are using.
Taking Offsets

Before data collection may begin, the electronic background signal should be measured.  This is accomplished by taking offsets.  The computer will then be able to subtract the background of electronic noise from the data during collection.  The procedure is both simple and brief:
 
Taking Offsets
1.  Close the shutters.
2.  Collect offsets.
 

Detuning

Synchrotron radiation may simultaneously satisfy the Bragg condition for multiple orders of diffraction (multiple n).  Bragg's Law is given by
 
nl = 2d sin q


where l is the wavelength, d is the lattice plane spacing (the d-spacing), and q is the Bragg angle.  The fundamental mode is such that n=1, and the harmonics are n such that n>1.  X-rays corresponding to harmonics (n>1) have energies that may interfere with data collection.  These should be removed prior to gathering data.  The angular widths of the diffraction conditions (rocking curves) for higher n are smaller than for the fundamental (n=1).  That is, reducing the parallelism of – or detuning – the crystals of the monochromator will significantly diminish the intensity of higher orders of diffraction, while having minimal impact upon the intensity of the first order.  Detuning causes the first monochromator crystal to be rotated off of the Bragg condition enough to greatly reduce the intensity of higher n, while maintaining much of the intensity of n=1.

How to Determine The Proper Amount of Detuning

To measure the amount of detuning required, a strong sample or foil of the element being investigated is placed in front of the Lytle detector.  The monochromator is then set to an energy below the binding energy of the element.  Thereby, most of the fluorescence emitted by the sample should result from harmonics in the beam, since they lie in higher energies than the fundamental.  Create a table similar to Table 5 that has the following columns: percent detuned, I0, IF, and IF/I0.  Make rows beginning with the monochromator fully tuned, 5% detuned, 10% detuned, and so on.  Record the values for I0 and IF in the table.  Then plot the ratio IF/I0 as a function of percent detuning.  When the ratio of IF/I0 exhibits asymptotic behavior as it approaches a constant value, the desired amount of detuning has been achieved.

As an example of the procedure for detuning, consider the actual detuning data collected for Uranium at beamline 4-3 of SSRL shown in Table 5 and Figure 5.  The ratio IF/I0 becomes asymptotic at 40%.  Therefore, as indicated in Table 5, the optimal detuning is 40%.  However, the experimenter should verify that this level of detuning is actually necessary by comparing the noise levels of scans with different amounts od detuning.

 
% Detuned at 17,100 eV
I0
IF
IF/I0
0% 2.5 .844 .34
5% 2.38 .818 .34
10% 2.25 .759 .34
15% 2.13 .695 .33
20% 2.00 .624 .31
25% 1.88 .568 .30
30% 1.75 .510 .29
35% 1.63 .458 .28
40% 1.50 .390 .26
45% 1.38 .364 .26
50% 1.25 .320 .26
 
 *  :closest-to-optimal detuning
Table 5.  Detuning Data
 Figure 5.  Detuning Graph
 


Once the amount of detuning is determined, the direction of detuning must also be determined: whether to detune up or down in Piezo voltage.  The direction of detuning can effect the severity of "glitches" (i.e. intensity spikes in I0) in data.  Detuning in one direction may reduce or even eliminate glitches present when tuned in the other direction.  A sample of the desired element should be run once with the monochromator tuned in one direction and once in the other.  Qualitative comparison will then yield the best direction to detune.  To measure glitches, determine the positions of the top and bottom glitch, which represent the counts per second of the data.  Then calculate its percent size.  It is undesirable to have glitches with a magnitude greater than 1%.
 

 


Before collecting data, the detuning should be checked at the lower and upper extremes of the energy region being scanned.  For example, suppose that Uranium is being scanned from 17,166eV to 18,040eV, and it is determined that 20% down is the optimal detuning.  The tuning is then set to 20% at 17,600eV.  The detuning is then checked at the extrema, with results summarized in Table 6.  At the upper end of the region, the tuning is only at 13%.  Therefore, the tuning should be adjusted as to insure that its value is 20% at the upper end.

 
Energy
% Detuning
17,166 25%
17,600 20%
18,040 13%
 Table 6.  Detuning Level at Selected Energies
 
    

Detuning should be checked often, and one should be mindful that the detuning may change as the temperature of the monochromator changes with variations in x-ray flux.