Stanford Synchrotron Radiation Laboratory
TXM SRL
  null   Depth of Focus

Along with the spatial resolution δ, the depth of focus (DOF) is a very important parameter for tomographic 3D imaging. The DOF is defined by DOF = 1.22 λ / (NA)2 = 3.36 δ2/λ, where λ is the wavelength and NA the numerical aperture of the objective zone plate lens. This expression shows that the DOF increases inversely with the wavelength, and thus linearly with the x-ray energy. Assuming a spatial resolution of 60 nm for first order diffraction, the DOF for a photon energy of 10 keV is about 100 µm.

  null   Downloads & Links

PDF iconIf you would like to read the full research paper please click on the PDF icon to download the research paper that explores the new TXM imaging technology at SSRL.