| CURRENT STATUS: |
Open |
| SUPPORTED TECHNIQUES: |
X-ray absorption spectroscopy imaging |
| MAIN SCIENTIFIC DISCIPLINES: |
Biology / Materials / Chemistry / Environmental |
| % TIME GENERAL USE: |
100% |
| SCHEDULING: |
Proposal Submittal and Scheduling Procedures Current SPEAR and Beam Line Schedules |
| SOURCE: |
1.3 Tesla Bend Magnet |
BEAM LINE SPECIFICATIONS:
| |
energy
range |
resolution DE/E
|
spot size |
flux |
angular acceptance
|
unfocused
|
4500-24000 eV
|
~5 x 10-4
|
2 x 2 microns
|
~1010
|
0.75 mrad
|
|
| OPTICS: |
Kirkpatrick-Baez system (microfocus) |
| MONOCHROMATOR: |
Si(111), Si(220) or Si(400), downward
reflecting, double-crystal
Monochromator Crystal Glitch Library
Crystal changes need to be scheduled and coordinated in advance with BL support staff. |
| ABSORPTION: |
584 microns Be, 10.7 meters He, 15.2 microns C |
| INSTRUMENTATION: |
Micro sample positioner
Ge3 detector
Vortex detector
CCD detector |
| DATA ACQUISITION AND ANALYSIS |
BL2-3 Start Up Guide X-ray Wiki Data aquisition: Microscan GUI
Data analysis: Microtoolkit
X-ray Data Booklet |
| RESPONSIBLE STAFF: |
Scientific Support: Sam Webb, 650-926-3734
Beam Line Support: Ben Kocar, 650-926-3769 Beam Line Technical Support: Bart Johnson, 650-926-3858
|
| BEAM LINE PHONE NUMBER: |
650-926-5223 On-site Users: Contact the Duty Operator at 9-926-4040 |
| GENERAL DESCRIPTION: |
BL2-3 is used a microfocus station used for micro x-ray absorption spectroscopy, micro x-ray absorption spectroscopy imaging, and micro diffraction. |
| Sample Preparation Labs: |
The SSRL Structural Molecular Biology (SMB) Sample Preparation Laboratories are available to users by prearrangement. To use a laboratory, users must complete laboratory training, and provide a list of all chemicals to be brought, used or stored in the Sample Preparation Laboratory. For details, contact Cynthia Patty (650-926-3925). |
SCIENTIFIC APPLICATIONS /
SELECTED RESULTS: |
|
| ENGINEERING NOTES: |
http://www-ssrl.slac.stanford.edu/safety/beamlines/bl2/ |